Avidity Science Geno II Type 2 pure water system

Avidity Science

Geno II

Geno II is a Type 2 pure water system that combines reverse osmosis and deionization in a single unit. It produces 25 or 45 liters per hour, delivers water of 1–15 MΩ·cm quality and is installed with a 30, 60 or 100 liter reservoir.

Request a quote
Water typeType 2 (deionized)
Production rate25 L/h (GENO20) or 45 L/h (GENO40)
Water quality1–15 MΩ·cm (25 °C)
Dispense rate3–7 L/min (from tank)
Reservoir30 · 60 · 100 liters

Geno II as a Type 2 pure water system

Geno II first passes mains water through a reverse osmosis membrane and then deionizes it with ion exchange resins. The product water meets the Type 2 quality defined for general laboratory use.

The system is designed for laboratories with a daily consumption of 160–320 liters. Water used for buffer and media preparation, sample dilution, reagent preparation, spectrophotometry, protein electrophoresis, cytology and histology falls into this class.

The unit can be mounted on the bench, on the wall or directly on top of the reservoir. Because a feed pump is built into the unit, no separate pump is required; this makes installation considerably easier in buildings with low mains pressure. For other options, see the Type 2 pure water systems section.

Consumables management and operating cost

The deionization pack, called Endure, is fitted and removed with a quick-connect system. The prefilter can also be replaced by the user. These two tasks, which require no service visit, account for most of the annual operating cost.

The system includes a recirculation process that extends the life of the RO membrane and the DI cartridge. In addition, an RO sanitization process prevents the buildup of microbial load.

  • TC001: Pretreatment module (on average 2 per year)
  • TC002: Endure purification pack 1 (1–2 per year)
  • TC005: Tank air filter (1 per year)
  • AV012: UV upgrade kit for tank (optional)
  • AV027: IQ/OQ documentation

All codes and the consumables for other systems are collected on the pure water cartridges and filters page.

Consumable requirements vary with feed water hardness and usage volume. When a mains water analysis report is shared, the annual consumables budget can be calculated realistically.

Installation requirements

A dedicated cold water line and a drain capacity of more than 3 liters per minute are required. The manufacturer specifies the following limits for feed water: conductivity of at most 1,400 µS/cm, hardness below 400 ppm, free chlorine below 2 ppm and a temperature of 4–30 °C.

The system measures 500 × 490 × 290 mm and weighs 23 kg when installed. The 30 and 60 liter reservoirs can be wall-mounted. The 100 liter reservoir with integrated pump can also supply a glassware washer and remote points of use.

Selection criteria for a Type 2 pure water system

If Type 1 ultrapure water is also needed, the Duo II·I combined system can be considered instead of two separate units; it produces both water types in a single unit and saves installation space. Alternatively, an Alto I Type 1 polishing system can be added to an existing Geno II, because Type 2 water is not sufficient for HPLC, cell culture and trace element analysis.

Choosing a reservoir volume above the daily consumption prevents the water from running out during peak hours. If consumption is not known, a cost analysis is carried out jointly at the quotation stage. Having information on daily usage, number of points of use and mains water ready at the first meeting therefore speeds up the process.

Technical specifications

Water typeType 2 (RO + deionization)
ModelsGENO20 (25 L/h) · GENO40 (45 L/h)
Water quality1–15 MΩ·cm (25 °C)
Inorganic removal98% (depending on feed water)
TOC reduction96% (with tank UV option)
Dispense rate3–7 L/min (depending on tank and pump setup)
Recommended usage160–320 L/day
Feed water0.1–6 bar · < 1,400 µS/cm · 4–30 °C · hardness < 400 ppm · free chlorine < 2 ppm · pH 4–10
Dimensions (H×W×D)500 × 490 × 290 mm
Installed weight23 kg
Electrical100–240 V, 50/60 Hz, max. 3.0 A
Reservoir options30 L · 60 L · 100 L · 100 L with pump
MountingBench, wall or on top of reservoir

Values are taken from the manufacturer's published technical documentation. They may vary by configuration and are confirmed against the current datasheet before quotation.

Frequently asked questions

For which tasks does Geno II provide sufficient water quality?

General laboratory tasks such as buffer and media preparation, dilution, reagent preparation, spectrophotometry, electrophoresis and glassware rinsing are performed with Type 2 water. HPLC mobile phase, cell culture and trace element analysis require Type 1 ultrapure water.

What is the difference between GENO20 and GENO40?

The difference is the production rate: GENO20 produces 25 liters per hour and GENO40 produces 45 liters per hour. The choice is based on daily consumption and reservoir volume; the recommended usage is in the 160–320 liters/day range.

Does Geno II require a separate pump for installation?

No. A feed pump is built into the unit, and the system can operate from a feed pressure of 0.1 bar. However, a dedicated cold water line and a drain capacity of more than 3 liters per minute must be provided.