The separation, drying, cryogenic freezing and measurement technologies behind the instruments supplied, which method is used in which application, and the related product groups.

Separation and purification

Flash chromatography

Flash chromatography is a fast liquid chromatography method used to isolate a target compound from a crude mixture after synthesis. The sample is passed through a disposable cartridge filled with packing material by means of a pressurized mobile phase. Compounds elute from the column at different times according to their affinity for the packing and are collected into separate tubes by the fraction collector.

The advantage of the method over classical column chromatography is speed and reproducibility. A process that takes hours on a manually run column is reduced to tens of minutes on an automated system, and because collection is triggered by the detector signal, the risk of fractions mixing is low. Resolution improves as particle size decreases; 15 µm packings deliver resolution close to that of preparative HPLC. Flash chromatography systems.

Preparative HPLC

Preparative HPLC applies the same separation principle as analytical HPLC with higher pressure and a larger column diameter. The aim is not to analyze but to collect the pure compound on a milligram to gram scale. High pressure allows the use of stationary phases with small particle sizes, which makes it possible to separate compounds that elute very close to one another.

The method is used in drug development for impurity isolation and reference standard production, and in natural product chemistry for collecting trace components. Preparative HPLC systems.

Pilot and process scale chromatography

When a purification method developed in the laboratory is transferred to production, the column diameter, flow rate and loading amount are scaled up proportionally. Pilot scale systems preserve the method during this transition and allow crude product to be processed semi-continuously on a kilogram scale. Pilot and process scale systems.

Analysis and detection

Mass spectrometry

A mass spectrometer measures molecules by ionizing them and separating them according to their mass-to-charge ratio. The result provides direct information on the identity of the compound. Compact mass spectrometers bring this technology from the central analytical unit to the chemist's bench, so that whether a reaction has produced the expected product is confirmed on the same day.

The choice of ion source determines the result. Electrospray ionization is preferred for polar and ionizable compounds, and atmospheric pressure chemical ionization for less polar compounds. Probe options that allow direct analysis of solid samples are also available. Mass spectrometry.

Thin-layer chromatography

TLC is the fastest and most economical method for reaction monitoring. The sample developed on a silica-coated plate is evaluated under UV light. Digital imaging of the plate allows the result to be recorded and reported.

Acquiring a mass spectrum directly from a spot on the TLC plate allows the identity of the compound to be confirmed without scraping and dissolution steps. TLC systems.

Gas chromatography

In gas chromatography, the sample is vaporized and passed through a capillary column together with the carrier gas. Separation takes place according to the affinity of the compounds for the stationary phase on the inner surface of the column. Phase chemistry, film thickness, column length and internal diameter are the four key parameters that determine the resolution and run time of the method.

Applications such as volatile organic compounds, residual solvents, pesticides and fatty acid methyl esters each require a different phase. GC columns.

Liquid chromatography

HPLC separates the components of a liquid sample according to differences in their distribution between the column packing and the mobile phase. Reversed phase columns cover most applications; normal phase, HILIC, ion exchange and size exclusion columns are used for specific separation requirements.

Particle size, pore size, pH stability and silica type are the decisive factors in column selection. Peak tailing observed with basic compounds often originates not from the mobile phase but from the silica surface of the column. HPLC / LC columns.

UV-Vis spectrophotometry

A UV-Vis spectrophotometer measures how much of the light passing through a sample is absorbed at specific wavelengths. Through the Beer-Lambert relationship, absorbance is converted into the concentration of the substance in solution; nucleic acid and protein quantification, kinetic studies and routine analyses based on color reactions are performed with this method. Surface color measurement, on the other hand, requires a different geometry and is performed with color measuring instruments. UV-Vis spectrophotometers.

Sample preparation and pure water

Pure water production

Laboratory water is defined in three grades according to its intended use. Type 3 water produced by reverse osmosis is used to feed autoclaves and glassware washers. Type 2 water, passed through ion exchange resin after reverse osmosis, is suitable for buffer preparation and general laboratory applications. Type 1 ultrapure water, with a resistivity of 18.2 MΩ·cm, low TOC and low bacteria levels, is used in HPLC, cell culture and trace element analysis.

The system is selected according to daily consumption, feed water quality and the number of points of use. The hardness and chlorine content of tap water have a direct effect on consumable life. Pure water systems · Combined water purification systems.

Evaporation

Solvent must be removed from the fractions obtained after purification. Parallel evaporation dries a large number of samples simultaneously using a heated nitrogen flow. The main difference from a rotary evaporator is that samples are processed in batches rather than one at a time, which eliminates the bottleneck that forms after purification. Evaporation systems.

Centrifugation

A centrifuge separates components of different density by centrifugal force. The decisive parameter of the method is not the rotational speed but the relative centrifugal force applied; at the same speed, rotors of different radii apply different forces. The rotor type also affects the result: a fixed-angle rotor provides fast pelleting, while a swing-out rotor provides clean phase separation. Centrifuge and vacuum systems.

Mixing and homogenization

Magnetic stirring is sufficient for low and medium viscosity liquids. When the volume increases or the mixture thickens, an overhead mechanical stirrer is used instead. In applications where the structure must be disrupted, such as tissue disruption and emulsion preparation, homogenizers operating on the rotor-stator principle are used. Mixing and shaking systems.

Heating and temperature control

Water baths and heating blocks are used in incubation, dissolution and reaction steps that require the sample to be held at a defined temperature. For reproducible results, the uniformity of temperature within the bath and its stability over time matter more than the setpoint. Heating and water bath systems.

Sample filtration and solid phase extraction

Removing particles from the sample before HPLC and GC analysis protects the column and the injection system. The syringe filter membrane is selected for compatibility with the solvent: PES or cellulose acetate is the common choice for aqueous samples and PTFE for organic solvents. For samples with high matrix interference, solid phase extraction (SPE) both cleans up the sample and increases analyte concentration. Syringe filters · SPE cartridges.

Rotary evaporation

A rotary evaporator heats the solution in a rotating flask under reduced pressure, evaporating the solvent below its boiling point; the vapor condenses in the condenser and is collected in a separate flask. For efficient operation, the vacuum pump, vacuum controller and recirculating chiller are planned together. Rotary evaporators and circulators.

Molecular biology

PCR and thermal cycling

PCR amplifies a target DNA region through temperature cycles consisting of denaturation, annealing and extension steps. The ability of the thermal cycler to change block temperature quickly and uniformly determines the reproducibility of the result; a gradient block allows the annealing temperature to be optimized in a single run. PCR thermal cyclers.

Electrophoresis

Electrophoresis separates charged molecules in a gel under an electric field according to their size. Nucleic acids are generally separated in horizontal agarose gels and proteins in vertical polyacrylamide gels (SDS-PAGE); the separated proteins are transferred to a membrane in a transfer tank and examined by western blot. Electrophoresis systems.

Drying and cryogenic freezing

Spray drying

In spray drying, a liquid sample is introduced into a hot air stream as fine droplets and turns into powder within seconds. Because the droplet spends a very short time in the hot zone, heat-sensitive active ingredients, flavors and probiotics can also be dried. The method is widely used at laboratory scale in pharmaceutical formulation, encapsulation and food ingredient development. Spray dryers.

Freezing with liquid nitrogen

Cryogenic freezing freezes the product with liquid nitrogen (LN2) or liquid carbon dioxide instead of a mechanical refrigeration unit. Liquid nitrogen boils at -196 °C at atmospheric pressure; cryogenic freezers operate at -100 °C and below, whereas mechanical freezers generally remain around -40 °C. The temperature difference between the product and the refrigerant is therefore more than twice as large, and freezing takes place much faster.

Advantages of using liquid nitrogen

  • Shorter freezing time: According to manufacturer data, the heat transfer coefficient is 100-140 W/m²·K in cryogenic systems and 15-17 W/m²·K in mechanical cold air systems; freezing time is reduced by an average of three quarters.
  • Smaller ice crystals: Rapid freezing forms ice crystals smaller than 10 µm inside the cells, whereas in mechanical freezing this size exceeds 50 µm. Small crystals preserve cell structure, so texture and appearance are not impaired after thawing.
  • Lower moisture and weight loss: Most of the loss occurs between -1 and -5 °C. Mechanical systems remain in this range for 10-30 minutes and lose 3-5 percent moisture; in cryogenic freezing this time is under 3 minutes and the loss is below 1 percent.
  • Smaller footprint: The same capacity can be achieved in an area up to 75 percent smaller than a mechanical system.
  • Hygiene: As there is no evaporator, fan or air duct in the product zone, cleaning is easier and the risk of contamination is reduced.
  • Control with biological samples: For samples such as cells, vaccines and gametes, the freezing rate can be programmed and reproducible profiles between -70 °C and -196 °C can be applied.

The cost structure also differs: cryogenic systems generally have lower capital cost and electricity consumption, but higher operating costs due to gas consumption. System selection should therefore be evaluated together with production volume and the quality requirements of the product. Controlled-rate freezers · Liquid nitrogen immersion and pelletizing systems · Cryogenic cabinet freezers · Industrial cryogenic freezing lines.

Measurement technologies

Viscosity measurement

Viscosity is the resistance of a liquid to flow. A rotational viscometer rotates a spindle immersed in the liquid at a constant speed and measures the resistance acting on the spindle. For non-Newtonian products, the result varies with rotational speed, so the measurement conditions must be stated in the report. The paint industry uses dedicated instruments that measure in Krebs units. Viscometers.

Color measurement

A spectrophotometer measures the amount of light reflected by the sample at each wavelength and converts the result into numerical coordinates in a color space. Differences between batches are thus expressed as numbers rather than by subjective visual assessment. Measurement geometry, aperture size and illuminant selection are parameters that affect the result. Gloss meters measure the specular reflection of a surface in GU at 20°, 60° or 85°; in applications where color and gloss must be captured in a single measurement, spectrophotometers with an integrated gloss sensor are used. Color and gloss measurement instruments.

Light and display measurement

An illuminance meter (lux meter) measures the amount of light falling on a surface in lux (lx). For workplace lighting inspections the illuminance level is sufficient, whereas in LED and lighting manufacturing the color of the light also becomes important. Color temperature (CCT) and chromaticity coordinates are determined with tristimulus light meters, while the color rendering index (CRI) and spectral distribution are determined with spectral illuminance meters. Cosine correction in the measuring head ensures that obliquely incident light is evaluated correctly. Light and display measurement instruments.

Moisture determination

A halogen moisture analyzer continuously monitors weight loss while heating the sample and calculates the moisture content within minutes. It is not expected to give exactly the same result as the drying oven method; the heating program of the instrument is calibrated once against the drying oven result, after which fast and consistent measurements are obtained. Balances and moisture analyzers.

pH and refractive index measurement

The accuracy of pH measurement depends largely on the condition of the electrode and on calibration; the meter only converts the signal from the electrode. A refractometer, in turn, determines the amount of dissolved substance by measuring the refraction angle of light passing through the sample. In both measurements, temperature control directly affects the result. pH and electrochemistry instruments · Refractometers.

The right solution, defined together

Based on the application and workload, suitable products and configurations are recommended with a clear rationale.