The DLAB heated stirring oil and water bath heats flasks and vessels uniformly in the bath while magnetically stirring both the bath fluid and the sample. The MS-H220-V3 and V5 models operate up to 220 °C; the TB-H240-V1 constant temperature bath offers PID-controlled heating up to 240 °C and a separate control unit.
Request a quote| Temperature | Room temperature to 220 °C (MS-H220) · 25–240 °C (TB-H240) |
|---|---|
| Stirring | 100–1,500 rpm (MS-H220) |
| Vessel | Borosilicate glass |
| Stirring capacity | 3 L (V3) · 5 L (V5) |
| Timer | 1 min to 99 h 59 min (MS-H220) |
Product description
An oil or water bath provides gentler and more uniform heat transfer than heating a flask directly on a hot plate. It is preferred especially in organic synthesis to prevent local overheating of the reaction flask and to keep the temperature within a narrow range.
The DLAB MS-H220-V3 and MS-H220-V5 models combine a hotplate stirrer with a glass bath vessel. While the bath fluid is stirred, the sample in the flask can also be stirred with a second magnetic stir bar, so both the bath and the sample heat uniformly. The TB-H240-V1 constant temperature bath reaches up to 240 °C and separates the control unit from the main body with a 1 meter cable, allowing the instrument to stay inside a fume hood while the controls remain outside.
Model comparison
| Model | Temperature range | Control accuracy | Vessel and capacity | Stirring | Power |
|---|---|---|---|---|---|
| MS-H220-V3 | Room temp. to 220 °C | ±1 °C | Ø220×105 mm, 3 L | 100–1,500 rpm | 850 W |
| MS-H220-V5 | Room temp. to 220 °C | ±1 °C | Ø240×135 mm, 5 L | 100–1,500 rpm | 1,300 W |
| TB-H240-V1 | 25–240 °C | ±0.5 °C | 150×100 mm glass vessel | With optional magnetic stirrer | 420 W |
The MS-H220 models feature a brushless DC motor, a "HOT" warning above 50 °C and 350 °C overheat protection. The TB-H240-V1 operates with a PT1000 external temperature sensor and 280 °C overheat protection; a protective cage surrounding the vessel prevents the vessel from tipping over and coming into contact with the heater.
Applications
- Controlled heating of flasks in a water or oil bath in organic synthesis
- Observation of the sample through the glass vessel during the reaction
- Dissolution and reflux setups
- Remotely controlled heating inside a fume hood (TB-H240-V1)
- University chemistry teaching laboratories
Correct use
The bath fluid level should be kept close to the sample level in the flask so that the sample heats evenly over its entire surface. Since water or moisture in an oil bath can cause spattering at high temperatures, the oil must be clean and dry. To monitor the temperature directly at the sample during heating, the PT1000 sensor should be positioned close to the flask. After work is finished, the instrument should be allowed to cool until the "HOT" warning goes out, and the vessel should be emptied only after that point.
Heated bath selection criteria
- Stirring: If the bath and the sample need to be stirred simultaneously, the MS-H220 models are selected.
- Volume: The V3 is suitable for stirring up to 3 L, and the V5 for larger baths and flasks.
- Temperature and control: If temperatures above 220 °C and a tighter tolerance are required, the TB-H240-V1 should be considered.
- Bath fluid: The flash point of oil used at high temperatures must be above the operating temperature.
For direct flask heating, stirring heating mantles, and for flat-bottomed vessels, MS-H series hotplate stirrers are alternatives. Suitable stir bars are listed on the magnetic stir bar set page, and other instruments are found in the heating category.
Technical specifications
| Heating range | Room temperature to 220 °C (MS-H220) · 25–240 °C (TB-H240-V1) |
|---|---|
| Control accuracy | ±1 °C (MS-H220) · ±0.5 °C (TB-H240-V1) |
| Display resolution | 0.1 °C |
| Vessel | Borosilicate glass, stainless steel base |
| Vessel size | Ø220×105 mm (V3) · Ø240×135 mm (V5) · 150×100 mm (TB-H240-V1) |
| Max. stirring volume (water) | 3 L (V3) · 5 L (V5) |
| Speed range | 100–1,500 rpm, ±10 rpm (MS-H220) |
| Max. stir bar length | 40 mm (MS-H220) |
| Timer | 1 min to 99 h 59 min (MS-H220) |
| Overheat protection | 350 °C (MS-H220) · 280 °C (TB-H240-V1) |
| External sensor | PT1000 (TB-H240-V1) |
| Protection class | IP21 |
| Power | 850 W (V3) · 1,300 W (V5) · 420 W (TB-H240-V1) |
| Power supply | 100–120 V / 200–240 V, 50/60 Hz |
Values are taken from the manufacturer's published technical documentation. They may vary by configuration and are confirmed against the current datasheet before quotation.
Frequently asked questions
When is an oil bath used instead of a water bath?
Since water boils at approximately 100 °C, an oil bath is required for work above this temperature. The MS-H220 models can heat up to 220 °C and the TB-H240-V1 up to 240 °C.
Can the bath and the sample be stirred at the same time?
Yes, on the MS-H220 models. The magnetic field that stirs the bath fluid also rotates the stir bar in the flask, so the bath and the sample heat uniformly together.
What is the advantage of the separate control unit on the TB-H240-V1?
The control unit can be placed up to 1 meter away from the main body. While the bath operates inside a fume hood, the controls remain outside and are not affected by vapors and contaminants.
From the same category

Stirring Heating Mantle
Digital heating mantles with magnetic stirring for 100–2,000 mL flasks, heating up to 280–380 °C.

Water Baths
Digitally controlled water baths; single and multi-compartment models.

Ultrasonic Baths
3–15 liter ultrasonic baths for cleaning, degassing and dissolution.

HB Series Dry Block Heater
Single or dual-module dry block heaters with LED display, operating from ambient +5 °C up to 150 °C.
Request a quote for this model
A suitable configuration and pricing are prepared based on sample type and workload.
Request a quote