MiniH, MiniHC and MiniHCL Mini Dry Block Heater
The DLAB MiniH series mini dry block heater takes up very little bench space with its housing weighing under 1 kg and heats up to 100 °C. The MiniHC and MiniHCL models also provide cooling down to 23 °C below ambient; the MiniHCL additionally prevents condensation on tube caps with its heated lid.
Request a quote| Temperature range | 0–100 °C (MiniHC, MiniHCL) · 25–100 °C (MiniH) |
|---|---|
| Control accuracy | ±0.5 °C |
| Heating rate | Up to 8 °C/min (MiniH: 6.5 °C/min) |
| Programs | 9 programs, 2 steps each |
| Weight | Under 1 kg |
Product description
A mini dry block heater is a palm-sized temperature control instrument designed for laboratories working with a small number of tubes and for field teams. With its 110 × 162 × 140 mm housing, the DLAB MiniH series fits easily on a bench, in a mobile testing vehicle or beside a fume hood.
The series consists of three models. The MiniH only heats; the MiniHC also provides cooling down to 23 °C below room temperature; the MiniHCL additionally comes with a heated lid. Because the instruments can be powered from a 100–240 V mains supply as well as from DC 12 V, they are suitable for in-vehicle and field use.
Model comparison
| Model | Temperature range | Heating rate | Cooling rate | Heated lid |
|---|---|---|---|---|
| MiniH 100 | 25–100 °C (from ambient +5 °C) | 6.5 °C/min | None | No |
| MiniHC 100 | 0–100 °C (down to ambient -23 °C) | 8 °C/min | 3 °C/min | No |
| MiniHCL 100 | 0–100 °C (down to ambient -23 °C) | 8 °C/min | 3 °C/min | Yes |
On all three models, heating from 25 °C to 100 °C takes no more than 20 minutes; the cooling models go from 20 °C to 0 °C in no more than 25 minutes. Temperature control accuracy is ±0.5 °C and display resolution is 0.1 °C.
Key features
- User temperature calibration maintains long-term accuracy
- 9 program memories; two steps can be defined in each program
- Time setting of 0–999 in minutes and seconds
- USB interface and error code display
- Interchangeable blocks for different tube sizes
- Overheat protection
The heated lid prevents condensation on tube caps during high-temperature steps. This reduces volume loss and concentration changes in small-volume reactions.
Applications
- Field and mobile molecular testing laboratories
- Preheating reagents and short-term storage at low temperature (MiniHC, MiniHCL)
- Nucleic acid and protein denaturation, enzymatic reactions
- Cell lysis and constant-temperature incubation in small batches
- Preparation of small numbers of samples in food and environmental analysis
Correct use
Contact between block and tube directly determines how temperature is transferred to the sample. A block matching the tube size should be used, and empty wells should be covered. On cooling models, condensation can form on the block surface if the environment is humid; after long low-temperature runs, the block should be dried before storage. Temperature verification can be carried out at regular intervals with a calibrated thermometer, and if a deviation is observed, the user calibration function is used.
Mini dry block heater selection criteria
If only heating is required, the MiniH is sufficient. If the sample must be kept below ambient temperature, the MiniHC is selected, and for protocols with small volumes and high-temperature steps, the MiniHCL.
For greater tube capacity and temperatures up to 150 °C, the HB series dry block heaters should be considered, and thermomixers for mixing during heating. All options are in the heating and water baths category.
Technical specifications
| Temperature setting range | Room temperature +5 °C to 100 °C (MiniH) · room temperature -23 °C to 100 °C (MiniHC, MiniHCL) |
|---|---|
| Control accuracy | ±0.5 °C |
| Display resolution | 0.1 °C |
| Heat-up time (25 °C → 100 °C) | Max. 20 min |
| Cool-down time (20 °C → 0 °C) | Max. 25 min (MiniHC, MiniHCL) |
| Time setting | 0–999 min / 0–999 s |
| Program memory | 9 programs, 2 steps each |
| Calibration | User temperature calibration |
| Interface | USB |
| Display | LCD |
| Power | 60 W |
| Power supply | 100–240 V, 50/60 Hz or DC 12 V |
| Dimensions (W × D × H) | 110 × 162 × 140 mm |
| Weight | Under 1 kg |
Values are taken from the manufacturer's published technical documentation. They may vary by configuration and are confirmed against the current datasheet before quotation.
Frequently asked questions
Can the mini dry block heater be used in a vehicle or in the field?
Yes. The series can run on DC 12 V as well as mains power and, weighing under 1 kg, is suitable for mobile testing vehicles and field laboratories.
How far can the MiniHC and MiniHCL cool?
Both models can go down to approximately 23 °C below ambient temperature; the operating range is specified as 0–100 °C. These models should be selected for steps where the sample must be kept cold.
What is the heated lid for?
The heated lid on the MiniHCL prevents water from condensing on the tube cap during high-temperature steps. It limits volume and concentration changes in small-volume reactions.
From the same category

Stirring Heating Mantle
Digital heating mantles with magnetic stirring for 100–2,000 mL flasks, heating up to 280–380 °C.

Water Baths
Digitally controlled water baths; single and multi-compartment models.

Ultrasonic Baths
3–15 liter ultrasonic baths for cleaning, degassing and dissolution.

HB Series Dry Block Heater
Single or dual-module dry block heaters with LED display, operating from ambient +5 °C up to 150 °C.
Request a quote for this model
A suitable configuration and pricing are prepared based on sample type and workload.
Request a quote