DUKO LifeDUKO Life
puriFlash 5.015-XL process scale flash system

Advion Interchim Scientific

puriFlash 5.015-XL

puriFlash 5.015-XL is a process scale flash system designed for high-volume and semi-continuous production. It offers 825 mL/min flow, column support up to F3000 format and a fraction collector that works with 29.5×200 mm tubes.

Request a quote
Flow / pressure825 mL/min · 15 bar
ScaleHundreds of grams to kilograms
ColumnF0040–F3000 flash
Fractions3 racks, 42 tubes (29.5×200 mm)
Display15-inch touchscreen

Product description

The 5.015-XL is a system designed to purify hundreds of grams of crude product in a single run and, under suitable conditions, quantities in the kilogram range. A pump flow of up to 825 mL/min is sufficient to run large F3000 format columns at their optimum flow rate.

The pressure limit is 15 bar; the model is designed for large-scale purifications carried out with silica of relatively coarse particle size. For stationary phases with small particle sizes that require high pressure, the 535-XL is more suitable.

The fraction collector works with 29.5×200 mm tubes. Being able to collect large-volume fractions without splitting them significantly simplifies the workflow in large-scale work.

Applications of the process scale flash system

It is suited to facilities that run semi-continuous production and purify the same product repeatedly.

  • Kilo-lab and process development units
  • Manufacturers purifying active pharmaceutical ingredient (API) intermediates in large batches
  • Facilities processing botanical extracts and natural products
  • Teams looking to transfer a method developed in R&D to production scale

Key features of the process scale flash system

Safety in semi-continuous production

Solvent and waste levels are monitored continuously; when a critical level is reached, the system pauses and the column is prevented from drying out. Leak detectors are fitted in the pump, UV detector and fraction collector compartments. These measures are intended to prevent product loss during long runs.

Pack-Process package

The package includes a column holder, a dry-load holder, a drip tray for dry-load cartridges, two RFID antennas and connection cables. If solid loading and automatic column recognition are to be used, including the package in the initial order is recommended.

Software

The system runs ClearPeak software with the Genius module. Features include the ITT function, which creates a purification method from TLC conditions, compound identification from a plate photo via the mobile app, and remote monitoring.

Pump

Seal wash discs are standard; they protect the pump heads against crystal buildup during long runs with mobile phases containing salts and additives.

Columns and complementary products

The 5.015-XL works with flash columns from F0040 to F3000 and dry-load cartridges up to F0330. Large batches typically use puriFlash silica flash columns, while reversed phase applications use reversed phase and functionalized columns. Dry-load cartridges and injection consumables are listed on the puriFlash columns and consumables page.

If pilot scale work with small-particle stationary phases is planned, the puriFlash 535-XL should be considered. At this scale, column selection directly determines solvent consumption and batch time, so the column format is planned together with the instrument. All models are listed in the pilot and process scale chromatography section.

Technical assessment

The manufacturer's technical sheet states column compatibility up to F1600, while the product page states up to F3000. The column format to be used is defined at the outset and confirmed in writing.

At this scale, infrastructure planning is as important as the instrument itself: solvent supply, waste collection, ventilation and power. A layout study is recommended before installation.

The instrument weighs 49 kg.

Technical specifications

Maximum flow825 mL/min (from 10 mL/min)
Maximum pressure15 bar
Flow accuracy±2% (100–750 mL/min, at 5–12 bar)
GradientQuaternary gradient valve
Pump protectionSeal wash discs
InjectionLiquid or solid; RFID dry-load recognition with Pack-Process
UV detector200–400 nm DAD; 200–800 nm with package
Flow cell0.3 mm / 80 µL
Column compatibilityF0040–F3000 flash (F1600 per technical sheet); dry-load up to F0330
Particle sizeNormal phase 15–50 µm · reversed phase 30–50 µm
Fraction collector3 racks · 42 tubes (29.5×200 mm), RFID rack recognition
Display and software15-inch touchscreen · ClearPeak + Genius · Windows 10
Connectivity8 USB, 2 RJ45, Bluetooth/Wi-Fi
SafetyLevel sensors, leak detectors, integrated fume enclosure
Power supply110 V or 220 V
Weight49 kg

Values are taken from the manufacturer's published technical documentation. They may vary by configuration and are confirmed against the current datasheet before quotation.

Frequently asked questions

At what scale is the 5.015-XL used?

It is designed to purify hundreds of grams of crude product in a single run and, under suitable conditions, quantities in the kilogram range. It is therefore preferred in facilities that run semi-continuous production and purify the same product repeatedly.

Why is there a 15 bar pressure limit?

The model focuses on processing large volumes at high flow and works with silica of relatively coarse particle size. Because these packings do not require high pressure, 15 bar is sufficient; for phases with small particle sizes, the 535-XL is more suitable.

What does the Pack-Process package include?

The package includes a column holder, a dry-load holder, a drip tray for dry-load cartridges, two RFID antennas and connection cables. Solid loading and automatic column recognition are used with this package.

Request a quote for this model

A suitable configuration and pricing are prepared based on sample type and workload.

Request a quote