Advion Interchim Scientific
puriFlash Silica Flash Columns
puriFlash silica flash columns provide normal phase purification with SI-HP, SI-HC and standard silica packings, 15–50 µm particle size and F0004–F1600 formats.
Request a quote| Phases | SI-HP, SI-HC, standard (irregular) silica |
|---|---|
| Particle size | 15, 25, 30, 50 µm (depending on phase) |
| Formats | F0004–F1600 |
| Pressure rating | 22 bar (ClearConnect: 15 bar) |
| Connection | Luer-lock inlet; compatible with any flash system |
Product description
Silica is the core packing for normal phase flash chromatography. puriFlash silica flash columns come in three silica grades to match the load and difficulty of the purification task. SI-HP is a high-performance silica preferred for difficult separations. SI-HC increases loading capacity through a higher surface area. Standard (irregular) silica is intended for routine, economical purifications.
The column body is made of clear polypropylene, with a Luer-lock inlet and a Luer-slip outlet. Interchim specifies these columns as compatible not only with puriFlash systems but with any flash system using standard connections. In the ClearConnect version, the outlet is also Luer-lock.
Which silica flash column for which task
- SI-HP (60 Å, 500 m²/g): Separation of closely eluting compounds; the 15 µm version delivers resolution approaching preparative HPLC
- SI-HC (60 Å, 680 m²/g): Tasks that require loading more sample on the same column size
- Standard silica (60 Å, 450 m²/g): Easy, frequently repeated routine purifications
- SI-HP AgNO3: Silver ion separation of isomers containing double bonds (e.g., fatty acids, terpenes)
If the sample does not dissolve in the mobile phase, a puriFlash dry-load cartridge is used with the column; the cartridge is also compatible with 15 µm columns.
Format selection
The format name indicates the approximate amount of silica in the column: F0004 contains approximately 4 g, F0040 approximately 40 g and F1600 approximately 1.6 kg of packing. Small formats suit reaction monitoring at milligram scale, while large formats suit purification at gram and kilogram scale. Sample load is generally planned at 1 to 10 percent of the packing mass, depending on the difficulty of the separation and the type of silica.
- F0004: 12 × 68 mm, 5–20 mL/min
- F0012 / F0025: 21 mm diameter, 15–50 mL/min
- F0040: 27 × 135 mm, 20–70 mL/min
- F0080 / F0120: 31–36 mm diameter, 30–150 mL/min
- F0220 / F0330: 60 mm diameter, 80–300 mL/min
- F0800 / F1600: 78–104 mm diameter, 150–500 mL/min
Technical evaluation
According to current Interchim information, standard puriFlash columns are stable up to 22 bar (319 psi) and ClearConnect columns up to 15 bar (218 psi). The high pressure rating allows operation at high flow rates with small particle size packings such as 15 µm. The columns are packed using Accurate Bed Technology; the consistent bed structure allows the same method to be reproduced across different lots.
Pore size, surface area and pH range are taken from Interchim catalogs and should be confirmed against the current data sheet for the phase and format to be ordered.
The columns work with flash chromatography systems such as the compact flash chromatography system puriFlash XS 530 and puriFlash 5.020. For polar compounds that are not retained on silica, reversed phase flash columns can be considered.
Technical specifications
| SI-HP | 60 Å, 500 m²/g; 15, 30, 50 µm; pH 1.5–6.5 |
|---|---|
| SI-HC | 60 Å, 680 m²/g; 15, 25, 50 µm; pH 1.5–6.5 |
| Standard silica | 60 Å, 450 m²/g; 20, 50 µm, irregular particles |
| SI-HP AgNO3 | 60 Å, 500 m²/g; 50 µm |
| Formats | F0004, F0012, F0025, F0040, F0080, F0120, F0220, F0330, F0800, F1600 |
| Pressure rating | Standard 22 bar / 319 psi; ClearConnect 15 bar / 218 psi |
| Connection | Luer-lock inlet, Luer-slip outlet (ClearConnect: Luer-lock outlet) |
| Body | Clear polypropylene |
| Compatibility | All flash systems with standard connections |
Values are taken from the manufacturer's published technical documentation. They may vary by configuration and are confirmed against the current datasheet before quotation.
Frequently asked questions
Can puriFlash columns be used on other brands of flash systems?
Yes. The columns have standard Luer connections and are specified by Interchim as compatible with all flash systems using standard connections. Connection compatibility is confirmed once the instrument model is provided.
How to choose between SI-HP and SI-HC?
SI-HP is preferred when the separation is difficult and the compounds elute close together; SI-HC is preferred when the separation is adequate but more sample needs to be loaded.
From the same category

puriFlash Columns and Consumables
Flash columns in more than 35 selectivities, dry-load cartridges and injection consumables.

puriFlash Reversed Phase and Functionalized Flash Columns
Prepacked flash columns in C18, CN, NH2, Diol, ion exchange, polymer and chiral phases; for polar and difficult separations.

puriFlash BIO Peptide and Biomolecule Columns
Reversed phase flash and preparative columns with 100, 200 and 300 Å pores; for synthetic peptide and biomolecule purification.

Preparative HPLC Columns (puriFlash PREP and Uptisphere)
Stainless steel preparative HPLC columns with 5, 10 and 15 µm packing, from 4.6 mm to 60 mm internal diameter, plus high-pressure dry-load.
Request a quote for this model
A suitable configuration and pricing are prepared based on sample type and workload.
Request a quote