puriFlash 535-XL is a pilot scale purification system designed for kilo-lab work and pilot production. Its dual-piston pump delivers 400 mL/min flow and 35 bar pressure; in addition to flash columns, it also works with dynamic axial compression (DAC) columns and preparative LC columns.
Request a quote| Flow / pressure | 400 mL/min · 35 bar |
|---|---|
| Scale | Up to several hundred grams per run |
| Column | F0004–F1600 flash; preparative and axial compression columns up to 80 mm ID |
| Fractions | 2 long racks, 112 tubes |
| Display | 15-inch touchscreen |
Product description
The 535-XL is a purification system used in the transition from laboratory scale to production scale. Its dual-piston pump delivers 400 mL/min flow at 35 bar pressure, so large columns can be run in a short time.
The 35 bar operating pressure is a significant advantage at pilot scale. Because stationary phases with 10 µm particle size in normal phase and 15 µm in reversed phase can be used, resolution and purity are maintained during scale-up. In low-pressure systems, purity usually decreases as the scale increases.
In addition to flash columns, dynamic axial compression (DAC) columns of any size and preparative LC columns can be connected. Both method development and production batches can be run on the same instrument.
Applications of pilot scale purification
It is suited to teams in the scale-up phase. Kilo-labs, process development and contract manufacturing laboratories are the main users.
- Chemical companies purifying active pharmaceutical ingredients (API) and intermediates
- Manufacturers processing cannabinoids, natural products and extracts
- Teams producing peptides and oligonucleotides
- Groups looking to transfer a method developed in R&D to production
Key features for pilot scale purification
Scale flexibility
Work from small flash columns to large preparative columns can be carried out on the same instrument. Once a method has been developed, it does not need to be redeveloped when the scale changes.
Column holder options
The colHolder package provides an integrated holder for columns in the F0012-F0330 range. For larger columns, an external high-capacity holder supporting up to F3000 format is available.
Software and monitoring
A 15-inch touchscreen, ClearPeak software, the Genius module and sample queue management are provided. Consecutive batches can be run, the process can be monitored remotely, and gradients can be created from TLC/HPLC data.
Durability
Solvent-wetted parts are made of stainless steel, Hastelloy, PTFE, PEEK and sapphire. The system is designed for continuous operation; wear parts are available as service parts.
Columns and scale-up
The 535-XL works with flash columns from F0004 to F1600, up to F3000 with the external holder, and with preparative and axial compression columns up to 80 mm internal diameter. puriFlash silica flash columns are used in normal phase, reversed phase and functionalized columns in reversed phase, and preparative HPLC columns for work requiring high resolution.
For teams working at laboratory scale with a flash system such as the puriFlash 5.020, the 535-XL is the scale-up step within the same product family. For larger batches, the puriFlash 5.015-XL can be considered; all models are listed in the pilot and process scale chromatography section.
Technical assessment
Column compatibility is stated differently in the manufacturer's documents: the technical sheet states compatibility up to F0330 for the integrated holder, while the product page states up to F3000 with the external holder. The column format to be used is confirmed in writing by the manufacturer before ordering.
No ELSD option is specified for this model. If compounds without chromophores make up most of the work, the 5.020 or 5.050 models, which offer an integrated ELSD package, should be considered.
The instrument weighs 36 kg. At pilot scale, solvent consumption and waste volume increase significantly; ventilation and waste management are planned before installation.
Technical specifications
| Maximum flow | 400 mL/min (from 10 mL/min) |
|---|---|
| Maximum pressure | 35 bar |
| Flow accuracy | ±2% (20–320 mL/min, at 5–28 bar) |
| Gradient | Quaternary gradient valve |
| Injection | Liquid or solid |
| UV detector | 200–400 nm DAD; 200–800 nm with package |
| Flow cell | 0.3 mm / 40 µL |
| Column compatibility | F0004–F1600 flash (up to F3000 with external holder); preparative and dynamic axial compression (DAC) columns up to 80 mm ID |
| Particle size | Normal phase 10–50 µm · reversed phase 15–50 µm |
| Fraction collector | 2 long racks · 112 tubes (18×150 mm) |
| Display and software | 15-inch touchscreen · ClearPeak + Genius · Windows 10 |
| Connectivity | 6 USB, 4 RS232, 1 RJ45, Bluetooth/Wi-Fi |
| Power supply | 110 V or 220 V |
| Weight | 36 kg |
Values are taken from the manufacturer's published technical documentation. They may vary by configuration and are confirmed against the current datasheet before quotation.
Frequently asked questions
How much product can the 535-XL purify in a single run?
Up to several hundred grams of crude product can be processed in a single run. The actual capacity varies with the column format, phase selection and difficulty of the separation.
What is a dynamic axial compression (DAC) column?
In these columns, the packed bed is compressed by a piston, producing a homogeneous bed even in large-diameter columns. The 535-XL works with these columns as well as with flash columns and preparative LC columns.
How to choose between the 535-XL and the 5.015-XL?
Because the 535-XL can use stationary phases with small particle sizes at 35 bar pressure, it maintains purity during scale-up. The 5.015-XL is limited to 15 bar, but with 825 mL/min flow it handles larger volumes and is preferred for large batches run with coarse-particle silica.
Request a quote for this model
A suitable configuration and pricing are prepared based on sample type and workload.