Konica Minolta CM-3700A spectrophotometer with transmittance measurement

Konica Minolta

CM-3700A Benchtop Spectrophotometer

The CM-3700A is a spectrophotometer that, in addition to d:8° reflectance measurement, measures transmittance with d:0° geometry; it achieves ΔE*ab 0.005 repeatability on the white plate. It offers three measurement apertures and computer-controlled continuous UV adjustment; its successor is the CM-3700A Plus.

Request a quote
Geometryd:8°, SCI or SCE; transmittance d:0°
RepeatabilityΔE*ab 0.005 (white plate)
Wavelength360–740 nm, 10 nm pitch
Measurement apertureLAV, MAV, SAV
Successor modelCM-3700A Plus

Product description

The CM-3700A is a benchtop spectrophotometer used for color measurements that require high accuracy. Reflectance is measured with d:8° geometry and can be switched between SCI and SCE; transparent samples are measured in transmittance mode with d:0° geometry.

Repeatability of ΔE*ab 0.005 on the white calibration plate and an average inter-instrument agreement of ΔE*ab 0.08 on 12 BCRA Series II tiles have allowed the instrument to serve as a reference in color standard management for many years.

The manufacturer states that the CM-3700A has been replaced by the CM-3700A Plus. The Plus model should be considered for new installations; compatibility, accessories and service for existing CM-3700A users are addressed separately.

Applications

  • Defining color standards and reference values
  • Color and color difference measurement of opaque samples in reflectance
  • Transmittance measurement of transparent sheets, films and liquids
  • Controlled definition of the UV condition for samples containing optical brighteners
  • Maintaining data continuity in existing CM-3700A installations

Technical evaluation of the transmittance spectrophotometer

The measurement area can be changed between LAV (Ø25.4 mm), MAV (Ø8 mm) and SAV (3×5 mm). For transmittance measurement, samples in sheet, plate or liquid form up to approximately 50 mm thick are accepted; sample length is not limited.

The UV content is adjusted continuously under computer control. Temperature drift is specified within ±0.10% per °C for spectral reflectance and ΔE*ab 0.05 per °C for color difference; it is therefore recommended to keep the measurement environment within 13–33 °C and as stable as possible.

The instrument connects to a computer via a USB 1.1 interface and is used with SpectraMagic NX or SpectraMagic NX2 software.

In transmittance measurement, the result depends directly on the sample thickness and the optical path length of the cuvette used. Measurements to be compared are therefore taken on sheets of the same thickness or with the same cuvette, and the cuvette surface is checked for fingerprints and scratches. In liquids, air bubbles and sediment also reduce transmittance. In addition, calibration is performed according to the measurement mode to be used.

Selection criteria

If sample temperature logging, the LMAV aperture and simultaneous SCI/SCE measurement are required, the CM-3700A Plus should be preferred. If 60° gloss is to be measured together with color, the CM-36dG series can be reviewed; for a benchtop instrument with a display that measures liquid and powder samples from above, the CM-5 can be considered.

Different geometry and aperture options can be compared in the color and gloss measuring instruments catalog. Color values are calculated according to the illuminant and observer conditions defined by the CIE colorimetry principles.

Technical specifications

Geometryd:8° (diffuse illumination, 8° viewing); di:8° SCI or de:8° SCE selectable
Transmittanced:0° (diffuse illumination, 0° viewing)
DetectorSilicon photodiode array, plane holographic grating
Wavelength range360–740 nm, 10 nm pitch, average half bandwidth approx. 14 nm
Measurement range0–200%, resolution 0.001%
Light sourcePulsed xenon arc lamp
Measurement time0.6–0.8 s (until start of data output)
Measurement / illumination areaLAV Ø25.4/28 mm; MAV Ø8/11 mm; SAV 3×5 / 5×7 mm; transmittance approx. Ø20 mm
RepeatabilityWhite plate: ΔE*ab 0.005, spectral 0.05%
Inter-instrument agreementAverage ΔE*ab 0.08; maximum ΔE*ab 0.3 on a single tile (LAV, 12 BCRA Series II tiles)
Temperature driftSpectral reflectance ±0.10%/°C; color difference ΔE*ab 0.05/°C
UV adjustmentComputer-controlled, continuous
Transmittance sampleSheet, plate or liquid; maximum thickness approx. 50 mm, unlimited length
InterfaceUSB 1.1
PowerAC 100–240 V, 50/60 Hz, 25 VA (with AC adapter)
Operating environment13–33 °C, 80% relative humidity or less, non-condensing
Dimensions / weight271 × 274 × 500 mm; 18 kg

Values are taken from the manufacturer's published technical documentation. They may vary by configuration and are confirmed against the current datasheet before quotation.

Frequently asked questions

Is the CM-3700A still available?

The manufacturer offers the CM-3700A Plus as the successor to the CM-3700A. Current availability and options for migrating to the Plus model are confirmed with the manufacturer at the quotation stage.

Can existing CM-3700A data be compared with the new model?

Both models measure with d:8° geometry over 360–740 nm. For data continuity, comparative measurement of the same samples on both instruments during the transition period is recommended.

What is UV adjustment used for?

The whiteness measured on paper, textiles and plastics containing optical brighteners depends on the UV content of the illumination. Continuous UV adjustment allows the measurement condition to be defined according to a specific standard illuminant.