Konica Minolta CM-3700A Plus reference spectrophotometer

Konica Minolta

CM-3700A Plus Reference Spectrophotometer

The CM-3700A Plus is a benchtop reference spectrophotometer designed for measurements that define color standards in reflectance and transmittance. It offers ΔE*ab 0.005 repeatability on the white tile, four measurement apertures, continuous UV adjustment and built-in sample temperature measurement.

Request a quote
Geometrydi:8° / de:8°; transmittance di:0° / de:0°
RepeatabilityΔE*ab 0.005 (white tile)
Inter-instrument agreementΔE*ab 0.08 (LAV/SCI)
Wavelength360–740 nm, 10 nm pitch
Sample temperatureRecorded with each measurement

Product description

The CM-3700A Plus is a benchtop spectrophotometer developed for reference measurements, where color standards are defined and other instruments are compared. Repeatability of ΔE*ab 0.005 on the white calibration plate and ΔE*ab 0.02 on the black tile makes the instrument suitable as the master measurement point in a supply chain.

Reflectance is measured in a Ø152 mm integrating sphere with d:8° geometry; SCI and SCE values can be acquired separately or simultaneously. The transmittance chamber accepts sheet, film and liquid samples up to approximately 50 mm thick; with the chamber cover open, sample length is not limited.

The main innovation that distinguishes the instrument from the previous generation is the recording of sample temperature during measurement. Because the color of some pigments and plastics changes with temperature (thermochromism), temperature data is important for interpreting measurement results.

Applications

The manufacturer specifies the instrument for use in the plastics, chemical, pharmaceutical, and paint and coatings industries.

  • Establishing corporate color standards and digital reference values
  • Formulation development and approval measurements in plastics and masterbatch production
  • Transmittance measurement of transparent liquids in the chemical and pharmaceutical industries
  • Color difference studies requiring high accuracy in paint and coatings R&D
  • Master reference station for comparing instruments at different sites

Key features

Four measurement apertures

The instrument can switch between SAV (3×5 mm), MAV (Ø8 mm), LMAV (Ø16 mm) and LAV (Ø25.4 mm) apertures. Samples ranging from small parts to large non-uniform surfaces are measured with the same instrument.

Continuous UV adjustment

The UV content is adjusted under computer control in 1,000 steps between 0% and 100%; a 400 nm UV cut filter is also provided. For paper, textiles and plastics containing optical brighteners, the illumination condition can be defined in accordance with standards.

Camera, WAA and ambient sensor

The built-in camera helps position the measurement spot; the image can be viewed in software such as SpectraMagic NX2. WAA technology allows spectral performance to be checked, while the ambient temperature sensor contributes to recording the measurement conditions.

Reference spectrophotometer selection criteria

The instrument is the successor to the CM-3700A; sample temperature measurement, the LMAV aperture and simultaneous SCI/SCE measurement have been added in the new model. If routine quality control and integrated gloss measurement take priority over reference-grade performance, the CM-36dG series is a more economical option. For measurements in the paper and pulp industry that require d:0° geometry, the CM-3630A should be considered.

Other models in the category are listed under color and gloss measuring instruments. Color values are calculated according to the illuminant and observer conditions defined by the CIE colorimetry principles.

Technical specifications

Reflectance geometrydi:8° / de:8°, SCI and SCE switchable and simultaneous; CIE No.15, ISO 7724/1, ASTM E1164, DIN 5033 Teil 7, JIS Z 8722 condition c
Transmittance geometrydi:0° / de:0°; JIS Z 8722 condition g
Integrating sphereØ152 mm
Detector38-element silicon photodiode array, diffraction grating
Wavelength range360–740 nm, 10 nm pitch, average half bandwidth approx. 14 nm
Measurement range0–200%, resolution 0.001%
Light sourcePulsed xenon arc lamp
Measurement / illumination areaSAV 3×5 / 5×7 mm; MAV Ø8/11 mm; LMAV Ø16/20 mm; LAV Ø25.4/28 mm; transmittance approx. Ø20/25 mm
RepeatabilityWhite plate: ΔE*ab 0.005, spectral 0.05%; black tile: ΔE*ab 0.02, spectral 0.02%
Inter-instrument agreementΔE*ab 0.08 (LAV/SCI, average of 12 BCRA Series II tiles)
UV adjustment400 nm UV cut filter; computer-controlled, 1,000 steps between 0–100%
Sample temperature accuracySAV ±1.2 °C; MAV and LMAV ±0.8 °C; LAV ±0.5 °C
Measurement timeSCI or SCE approx. 2 s; SCI+SCE approx. 5 s; transmittance approx. 2 s
Transmittance chamberMaximum sample thickness approx. 50 mm; unlimited sample length with cover open
OtherBuilt-in camera, WAA internal performance check, ambient temperature sensor, USB 2.0
Operating environment13–33 °C, 80% relative humidity or less, non-condensing
Dimensions / weight307 (H) × 271 (W) × 600 (D) mm; approx. 20.0 kg

Values are taken from the manufacturer's published technical documentation. They may vary by configuration and are confirmed against the current datasheet before quotation.

Frequently asked questions

What is the difference between the CM-3700A Plus and the CM-3700A?

The Plus model records sample temperature with each measurement, adds the LMAV (Ø16 mm) aperture, can measure SCI and SCE simultaneously, includes a camera and WAA performance check, and operates with a USB 2.0 interface.

Why is measuring sample temperature important?

The color of some pigments, paints and plastics changes with temperature. Recording the temperature at the time of measurement allows results taken at different times or in different laboratories to be compared correctly.

Which samples are measured in transmittance mode?

Transparent sheets, films and liquid samples are measured in the transmittance chamber. Glass and plastic cuvettes with 2, 10 and 20 mm optical path lengths are offered as optional accessories for liquids.

When is a reference spectrophotometer preferred?

It is preferred where the color standard is defined on a master instrument and distributed to instruments at suppliers and sites, and where measurement uncertainty must be kept to a minimum.