DLAB
DM0 Series Low-Speed Centrifuge
The DM0 series low-speed centrifuge consists of benchtop models developed for serum and plasma separation, urine processing and pre-preparation of clinical samples. The series includes eight models, from the compact 6-tube DM0306 to the DM0636 with 4 × 100 mL capacity; the DM0612 will be available soon.
Request a quote| Maximum speed | 3,400–6,500 rpm (depending on model) |
|---|---|
| Maximum RCF | 1,600×g to 4,100×g (depending on model) |
| Capacity | From 6 × 3–10 mL blood tubes to 4 × 100 mL |
| Preset programs | Blood, urine, stool (DM0408, DM0424) |
| Motor | Brushless DC (except DM0306) |
Product description
The low-speed centrifuge, working with blood collection tubes and 10–50 mL conical tubes, is the most frequently used instrument in clinical and routine laboratories. Speeds of a few thousand rpm are sufficient for separating serum and plasma from cells, sedimenting urine and harvesting cell cultures.
Most models in the series run on a maintenance-free brushless DC motor, and users can switch between speed and RCF. The lid lock opens automatically at the end of the run. The DM0408 and DM0424 provide preset programs for blood (3,200 rpm, 10 min), urine (1,800 rpm, 5 min) and stool (1,300 rpm, 10 min); the DM0506 starts two programs with a single key.
The DM0306 is the most compact model and operates at 1,600×g with 6 blood collection tubes; its transparent lid allows the run to be observed. The DM0636, with 4 × 100 mL capacity, 10 program memories and automatic rotor recognition, is the largest model in the series.
Model comparison
| Model | Speed range | Max. RCF | Rotor capacity | Display | Weight |
|---|---|---|---|---|---|
| DM0306 | 3,400 rpm | 1,600×g | 6 × 3–10 mL blood collection tubes | LED | 5.3 kg |
| DM0408 | 300–4,000 rpm | 1,900×g | 8 × 15 mL or 12 × 10 mL (angle) | LED | 9.2 kg |
| DM0412 | 300–4,500 rpm | 2,490×g (A12-10P) · 1,800×g (A6-50P) | 12 × 1.5–10 mL · 6 × 15/50 mL | LCD | 6 kg |
| DM0424 | 500–4,000 rpm | 2,500×g | Swing-out 6 × 15 mL or 4 × 50 mL · angle 24 × 15 mL | LCD | 13.2 kg |
| DM0436E | 100–3,500 rpm | 2,260×g | 12 × 15 mL + 12 × 10 mL | LCD | 15 kg |
| DM0506 | 300–5,000 rpm | 2,600×g | 6 × 1.5–15 mL | LCD | 5.2 kg |
| DM0636 | 300–6,000 rpm | 4,100×g | 4 × 100 mL | LCD | 36 kg |
| DM0612 (coming soon) | 300–6,500 rpm | 4,000×g | 6 × 60 mL (angle) | LCD touchscreen | 12 kg |
Applications
- Serum, plasma and whole blood separation
- Processing of urine and stool samples
- Pre-analytical preparation of clinical samples
- Cell culture harvesting, microorganism pelleting and 96-well plate centrifugation (DM0636)
The typical workflow in clinical laboratories is described on the applications page.
Low-speed centrifuge selection criteria
- Tube type: For blood collection tubes, the DM0306 or DM0436E is suitable; for 15 and 50 mL conical tubes, the DM0412 with the A6-50P rotor or the DM0424 with a swing-out rotor is suitable.
- Rotor type: A swing-out rotor provides a clean phase separation in the horizontal position; an angle rotor speeds up pelleting.
- Daily sample count: For processing up to 24 tubes at once, the DM0424 or DM0436E is considered; for large volumes, the DM0636.
- Program requirement: If different users need to apply the same protocol, models with preset programs or program memory are preferred.
15 and 50 mL conical centrifuge tubes can be used with the series. For work with microtubes that requires high force, a high-speed microcentrifuge, and for higher speed and capacity, a high-speed benchtop centrifuge can be reviewed. All options are found in the centrifuges and vacuum category.
Technical specifications
| Speed range | From 3,400 rpm (DM0306) to 300–6,500 rpm (DM0612) |
|---|---|
| Maximum RCF | From 1,600×g (DM0306) to 4,100×g (DM0636) |
| Speed accuracy | ±20 rpm (DM0412, DM0436E, DM0506, DM0612, DM0636) · ±100 rpm (DM0408, DM0424) |
| Run time | From 30 s to 30 min up to 30 s to 99 h depending on model; continuous operation |
| Acceleration / deceleration | 1 to 9 acceleration levels and 2 to 10 deceleration levels depending on model |
| Programs | Blood, urine, stool preset programs (DM0408, DM0424) · P1/P2 (DM0506) · 9 programs (DM0612) · 10 programs (DM0636) |
| Motor | Brushless DC (AC fixed-frequency brushless motor on the DM0306) |
| Display | LED (DM0306, DM0408) · LCD (others) · LCD touchscreen (DM0612) |
| Safety | Lid lock, overspeed detection, automatic self-diagnosis (depending on model) |
| Power supply | Single phase 100–240 V, 50/60 Hz (depending on model) |
| Weight | From 5.2 kg (DM0506) to 36 kg (DM0636) |
Values are taken from the manufacturer's published technical documentation. They may vary by configuration and are confirmed against the current datasheet before quotation.
Frequently asked questions
What speed is required to separate serum?
Serum and plasma separation is usually performed with a low-speed centrifuge; the preset blood program on the DM0408 and DM0424 is defined as 10 minutes at 3,200 rpm. The force and time recommended by the manufacturer of the tube used should be followed.
What is the difference between a swing-out rotor and an angle rotor?
In a swing-out rotor, the tubes swing to a horizontal position during the run and the phase boundary forms flat; in an angle rotor, the tubes remain at a fixed angle and the pellet is collected faster.
When will the DM0612 be available?
The DM0612 is listed by the manufacturer as coming soon. Preliminary information and the delivery schedule are shared at the quotation stage.
From the same category

High-Speed Microcentrifuge (D1524, D2012, D3024)
High-speed microcentrifuges reaching 15,000 rpm and 21,380×g for nucleic acid and plasmid work.

D1008 and D1012UA Mini Centrifuge
Palm-sized 7,000 and 10,000 rpm mini centrifuges for quick spin-down of PCR tubes and microtubes.

D1524R and DG1616R Refrigerated Centrifuge
15,000–16,000 rpm refrigerated centrifuges cooling down to -20 °C for heat-sensitive samples.

DM1224, DM1424 and DG1616 High-Speed Centrifuge
12,000 rpm DM series for capillary tube applications and 16,000 rpm DG1616 with up to 4 × 400 mL capacity.
Request a quote for this model
A suitable configuration and pricing are prepared based on sample type and workload.
Request a quote