DUKO LifeDUKO Life
puriFlash 5.050 flash and preparative purification system

Advion Interchim Scientific

puriFlash 5.050

puriFlash 5.050 is a system that combines flash and preparative purification in a single housing. It offers 50 bar pressure, compatibility with preparative columns up to 80 mm internal diameter, and UV, ELSD and MS detection; it is preferred for difficult separations and work requiring high purity.

Request a quote
Flow / pressure250 mL/min · 50 bar
DetectionUV 200–400 nm; 200–800 nm and integrated ELSD with packages
Column rangeF0004–F3000 flash; preparative up to 80 mm ID
Fractions3 racks, 132 tubes (18×150 mm)
Injection2 mL steel loop, 6-way valve

Product description

The 5.050 is a model that bridges the gap between flash chromatography and preparative HPLC. Its 50 bar operating pressure allows the use of stainless steel preparative columns and reversed phase silicas with small particle sizes. The same instrument also continues to work with disposable flash columns.

Reusable columns lower consumable costs over the long term; in laboratories that use hundreds of columns per year, the difference is significant. Reduced solvent consumption also lowers waste costs.

Seal wash discs are located behind the pump heads. When working with buffers and additives that can crystallize, water is circulated behind the pistons to prevent crystal buildup; this design extends pump life.

Applications of flash and preparative purification

It is suited to laboratories where flash chromatography falls short but a fully equipped preparative HPLC system is not needed.

  • Medicinal chemistry groups performing difficult reversed phase separations
  • Metabolite isolation and natural product chemistry work
  • Teams purifying peptides and oligonucleotides
  • High-throughput laboratories looking to reduce the cost of disposable columns

If only routine flash purification is planned, this model may be more than required. The 5.050 is recommended for difficult separations and for applications where a move to preparative columns is planned.

Key features for flash and preparative purification

Injection flexibility

An RFID-tagged 2 mL stainless steel injection loop is standard. With the 6-way automatic valve, both loop injection and solid loading (dry-load) can be performed. The installed loop and cartridge are identified automatically and recorded in the method.

Triple detection

Detection is performed by UV (200-400 nm; 200-800 nm with package), integrated ELSD and mass spectrometry (MS). When the three detectors are used together, all fractions, including impurities without chromophores, can be monitored.

Column compatibility

Flash columns from F0004 to F3000 and preparative columns up to 80 mm (3 inch) internal diameter can be used. Particle sizes down to 10 µm in normal phase and 15 µm in reversed phase are supported.

Protection

Separate leak detectors are fitted in the pump, dry-load holder, UV detector, valve, fraction collector and ELSD compartments. This design is intended to prevent the loss of valuable products obtained at the end of multi-step syntheses.

Columns and complementary systems

In addition to flash columns from F0004 to F3000, the 5.050 works with stainless steel preparative columns up to 80 mm internal diameter. On the flash side, silica flash columns and reversed phase and functionalized columns are used; on the preparative side, preparative HPLC columns. For peptide purification, wide-pore puriFlash BIO columns offer a separate option.

For MS detection, the expression CMS mass spectrometer can be connected to the system. The other models in the series are listed in the flash chromatography systems and preparative HPLC systems sections.

Technical assessment

If preparative columns are to be used, the column, stationary phase and fittings should be planned together with the instrument; a first-year consumables list is provided with the quotation.

The instrument weighs 49 kg. The manufacturer refers to the software as ClearPeak in some documents and as InterSoft X in others; the version to be delivered is confirmed in writing at the ordering stage.

Technical specifications

Maximum flow250 mL/min (from 1 mL/min)
Maximum pressure50 bar
Flow accuracy±2% (20–200 mL/min, at 5–40 bar)
GradientQuaternary gradient valve; gradient step accuracy < 3.5%
Pump protectionPump seal wash discs
InjectionLiquid injection or solid loading; 6-way automatic valve, RFID-tagged 2 mL stainless steel loop
UV detector200–400 nm DAD; 200–800 nm with package
ELSDIntegrated with iELSD package
Flow cell0.1 mm / 40 µL
Column compatibilityF0004–F3000 flash; preparative up to 80 mm ID
Particle sizeNormal phase 10–50 µm · reversed phase 15–50 µm
Fraction collector3 racks · 132 tubes (18×150 mm), RFID rack recognition
Display and software15-inch touchscreen · ClearPeak / InterSoft X + Genius
Connectivity8 USB, 2 RJ45, Bluetooth/Wi-Fi
Power supply110 V or 220 V
Weight49 kg

Values are taken from the manufacturer's published technical documentation. They may vary by configuration and are confirmed against the current datasheet before quotation.

Frequently asked questions

What is the difference between the 5.050 and the 5.250?

The 5.050 operates at 50 bar and combines flash and preparative purification in one instrument. The puriFlash 5.250, with 250 bar pressure, goes down to columns with 5 µm particle size and is technically a preparative HPLC system.

What is the advantage of a reusable column?

Stainless steel preparative columns have a long service life, so consumable costs drop significantly in laboratories that use many columns per year. Reduced solvent consumption also lowers waste costs.

What are the seal wash discs for?

The seal wash discs behind the pump heads circulate water behind the pistons when working with buffers and additives. This prevents crystal buildup and extends pump life.

Request a quote for this model

A suitable configuration and pricing are prepared based on sample type and workload.

Request a quote