Uni Gloss 60A Single-Angle Gloss Meter

Uni Gloss 60A is a compact, single-angle gloss meter with a 60° measurement angle, designed for routine quality control. It provides statistics, difference and pass/fail evaluation through a simple two-button interface.

Request a quote
Measurement angle60°
Measurement range0–1000 GU
Measurement area9×15 mm
Memory999 measurements, 50 standards
PowerAA alkaline battery or USB

Product description

Uni Gloss 60A is a single-angle gloss meter developed for the daily inspection of medium-gloss surfaces. The 60° geometry used by the instrument is the reference angle defined in ISO, ASTM, DIN, BS and JIS standards; it is also the most widely used gloss measurement in the paint and plastics industries.

Language setting, calibration, measurement and statistics are all operated with two buttons. An automatic diagnostic function monitors the instrument status and alerts the user when calibration is required. When not in use, the instrument is protected by placing it in its calibration holder.

Applications

  • Routine gloss control of paint and plastic products
  • Incoming inspection and production control of building materials
  • Fast pass/fail decisions on the production line
  • On-site gloss verification after coating

Technical evaluation of a single-angle gloss meter

In the 0-99.9 GU part of the measurement range, resolution is 0.1 GU, repeatability 0.2 GU and reproducibility 0.5 GU. At higher values, i.e. 100 GU and above, resolution is 1 GU; repeatability is 0.2% of reading and reproducibility 0.5% of reading.

The instrument stores up to 999 measurements for statistical calculation and can save 50 standards. It is powered by an AA alkaline battery or via the USB port. Turkish is among the available interface languages.

A single-angle instrument is sufficient for operations whose product range is concentrated in the medium-gloss band. However, if the value measured at 60° is very high or very low, the surface lies outside the range in which this angle is most sensitive; in that case an instrument capable of measuring at 20° or 85° is required.

When the pass/fail function is used, the reference value and tolerance are defined on an approved standard sample. The operator on the production line then only reads the result, and the decision requires no separate calculation. The calibration standard is stored with the instrument, its surface is kept clean and the check intervals are written into the quality plan.

Because the instrument runs on an AA alkaline battery, no charging infrastructure is needed in the field; keeping a spare battery prevents interruptions to the measurement plan. In addition, since recorded measurements are kept in the instrument for statistics, mean and deviation values can be read directly from the display at the end of a shift.

Selection criteria

If Bluetooth connectivity and a color display are required, the Novo-Gloss 60 KM is an alternative. If matte and high-gloss surfaces are also to be measured, the three-angle Multi Gloss 268A should be preferred.

Different geometry and aperture options can be compared in the color and gloss measuring instruments catalog. The method for specular gloss measurement is defined in the ASTM D523 standard.

Technical specifications

Measurement angle60°
Measurement area9×15 mm
Measurement range0–1000 GU
Resolution0–99.9 GU: 0.1 GU · 100 GU and above: 1 GU
Repeatability0–99.9 GU: 0.2 GU · 100 GU and above: 0.2% of reading
Reproducibility0–99.9 GU: 0.5 GU · 100 GU and above: 0.5% of reading
StandardsISO, ASTM, DIN, BS and JIS gloss standards
FunctionsStatistics, difference measurement, pass/fail, automatic diagnostics
Memory999 measurements, 50 standards
Power1.5 V AA (R6) alkaline battery or USB
Interface languagesEnglish, Spanish, German, French, Italian, Japanese, Portuguese, Russian, Turkish, Polish

Values are taken from the manufacturer's published technical documentation. They may vary by configuration and are confirmed against the current datasheet before quotation.

Frequently asked questions

How does the pass/fail function work?

Once a reference value and tolerance have been defined, each measurement is compared with these limits and the result is shown on the display as pass or fail. This enables fast decisions on the production line.

Which gloss level is the Uni Gloss 60A suitable for?

The instrument is recommended for medium-gloss samples. For surfaces reading above 70 GU or below 10 GU at 60°, a three-angle instrument that also offers the 20° or 85° angles provides more accurate differentiation.

What is the difference between the 20°, 60° and 85° measurement angles?

The 60° angle is the reference geometry for medium-gloss surfaces and serves as the starting point for most measurements. On high-gloss surfaces reading above 70 GU at 60°, 20° is preferred to increase differentiation, while 85° is preferred for matte surfaces below 10 GU. This classification is consistent with the approach in ISO 2813 and ASTM D523.